INPUT CONNECTORS
03.10
01/2011 We reserve the right to make technical changes.
0033
www.ahlborn.com
ALMEMO
®
Connectors for DC Current mA
Types:
Model Meas. Range Resolution
32mA DC –32.0 to +32.0 * 1μA Order no. ZA9601FS1
4/20mA DC 0 to 100% 0.01% Order no. ZA9601FS2
2 x 32 mA DC –32.0 to +32.0 * 1 μA no electrical isolation Order no. ZA9601FS3
2 x 4/20 mA DC 0 to 100% 0.01 % no electrical isolation Order no. ZA9601FS4
*Data may vary depending on device; (see data sheet per device).
ALMEMO
®
Connectors for DC mA Differential
for sensors / transmitters, Supply from the ALMEMO
®
device
Types:
Model Meas. Range Resolution
32 mA DC –32.0 to +32.0 * 1 μA Order no. ZA9601FS5
4/20 mA DC 0 to 100% 0.01% Order no. ZA9601FS6
*Data may vary depending on device; (see data sheet per device)
ALMEMO
®
Connectors for DC mA Differential
for sensors / transmitters, Supply : 12 V from the ALMEMO
®
device
Types:
Model Meas. Range Resolution
32mA DC –32.0 to +32.0 * 1 μA ZA9601FS5V12
4-20mA DC 0 to 100% 0.01 % ZA9601FS6V12
*Data may vary depending on device; (see data sheet per device).
Technical Data:
Sensor supply U
F
: 13.5V ±0.5V
Output current: 100mA at U
G
= 12V
50mA at U
G
= 9V
20mA at U
G
= 7V
Accuracy shunt: ±0,1% of measured value
(U
G
= device voltage)
Technical Data:
Accuracy shunt: ±0,1% of measured value
Technical Data:
Sensor supply (for voltage see technical data
of ALMEMO
®
device)
Accuracy shunt: ±0,1% of measured value
2-wire transmitter
(3-wire transmitter can also be
connected)