Pressure Modules for Dynamic Pressure
Example:
Air velocity 50 m/s, air temperature 80 °C, atmospheric pressure 960 mbar.
The measured value must be multiplied with the correction value 1.123. The air
velocity is, therefore, 56.1 m/s.
ALMEMO
®
Pressure Connector:
For flow measurements the ALMEMO
®
sensor range provides plug-on pres-
sure connector FDA602-SxK and, as accessories, robust stainless steel or
nickel-plated brass Prandtl Pitot tubes. They are connected to the pressure
connector via hoses. Cylindrical probes can be used if the specific probe coeffi-
cient (1.7) is considered by programming the factor 1/
= 0.767.
The measuring variables, dynamic pressure and air velocity are programmed
within the ALMEMO
®
sensor connector on two measuring channels and can be
activated and displayed with the correct scaling and dimension by every AL-
MEMO
®
measuring instrument.
Pressure Connector FDA602-SxK Pitot tube FD9912
Designation Channel Meas. Range Dim Range Factor Exp
FDA602-S1K: 1st chan: 0.5...40.0 m/s ms L840 - -
2nd chan: ± 1250.0 Pa Pa Volt - 3
FDA602-S6K: 1st chan: 1.8...90.0 m/s ms L890 - -
2nd chan: ± 6800 Pa Pa Volt 0.4 4
3-5-16 ALMEMO
®